8-6
<EMISSIONS>
<IMMUNITY>
Table 1ENCLOSURE PORT
Test
Standard
Compliance
HOME HEALTHCARE
ENVIRONMENT
Conducted Emissions
CISPR 11
Group 1 Class B
Radiated Emissions
CISPR 11
Group 1 Class B
Harmonic distortion
IEC 61000-3-2
Harmonic Class A
Voltage fluctuations and
flicker
IEC 61000-3-3
Complies
Test
Standard
IMMUNITY TEST LEVELS
HOME HEALTHCARE
ENVIRONMENT
ELECTROSTATIC
DISCHARGE
IEC 61000-4-2
8 kV contact
2
kV
4
kV
8 kV,
15
kV
air
Radiated RF EM
fields
IEC 61000-4-3
10
V/m
80MHz 2.7GHz
80% AM at 1kHz
Proximity fields from
RF wireless
communications
equipment
IEC 61000-4-3
See Table 2
RATED
power
frequency
magnetic
fields
IEC 61000-4
-
8
30 A / m
50Hz or 60Hz