8-8
Table 3Input a.c. power PORT
Table 4Signal input / output parts PORT
Test
Standard
IMMUNITY TEST LEVELS
HOME HEALTHCARE ENVIRONMENT
Electrical fast
transients / b
ursts
IEC 61000-4-4
2
kV
100kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
0.5 kV
1 kV
Surges
Line-to-ground
IEC 61000-4-5
0.5 kV
1kV
2kV
Conducted
disturbances
induced
by RF fields
IEC 61000-4-6
3
V
0.
15MHz - 80MHz
6V in ISM and amateur radio bands between
0.
15MHz and 80MHz
80% AM at 1kHz
Voltage dips
IEC 61000-4-11
0%
UT
;
0.
5 cycle
At
0
°
,
45°,
90
°
,
135
°
,
180
°
,
225°
,
270°
and
315°
0%
UT; 1 cycle
and
70
%
UT; 25 / 30 cycles
Single phase: at 0°
Voltage interruptions
IEC 61000-4-11
0%
UT
; 250 / 300 cycle
Test
Standard
IMMUNITY TEST LEVELS
HOME HEALTHCARE ENVIRONMENT
ELECTROSTATIC
DISCHARGE
IEC 61000-4-2
8 kV contact
2 kV
4 kV
8 kV
15 kV air
Electrical fast
transients / bursts
IEC 61000-4-4
1 kV
100 kHz repetition frequency
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3
V
0.15 MHz - 80 MHz
6 V in ISM and amateur radio bands between
0.15 MHz and 80 MHz
80 % AM at
1 kHz