Operation Manual
20
4.11.2. Executing user’s program
To execute user-defined program stored in device memory simply press F1’, F2
or ‘F3’ button. Chart Panel will automatically display first test chart of the program
you selected.
In the left bottom corner of the screen a letter ‘A’ is displayed to inform that user
program mode is executed.
Use right Arrow to move to the next chart of the program, left Arrow to return to
the previous chart.
To switch to other program or to restart selected program press respectively F1’,
F2’ or ‘F3’ button.
To enter normal operating mode press any other button than F1’, ‘F2’, ‘F3’.
4.12. Auto-Off
This function allows user to preset time after which Chart Panel enters stand-by
mode. You can set this time from 4 to 10 minutes.
1. Enter ‘Auto-Offusing Arrow Up or Arrow Down button.
2. Press Right Arrow to edit time for automatic auto off.
3. Use Up and Down Arrows to set required time.
4. To store selected value press Left Arrow.
Power OFF – new power saving feature.
TCP-2000 Chart Panel will turn itself off when not used after preset time. This
time is set in factory to 4 hours. Service personnel can adjust it during installation
to 30min, 1h, 2h, 3h and 4h depending on the customer needs. If TCP-2000 is in
off-mode LED on the front panel is also off.
To restart TCP-2000 simply press any key on the remote control. Time needed for
startup is approximately 45sec.
NOTE !