3-20
If no readings have been stored, nothing will be displayed
when the READING button is pressed.
To return to the measurement display, remove the lens
from the nose piece and either place a lens on the nose
piece again or press any button, except the READING
button.
Previously stored readings will stay in memory until new
readings are stored , either by pushing the HOLD button in
MANUAL mode or automatically in AUTO mode.
3.2.7 DETECTING PROGRESSIVE LENSES
When the instrument is in single Vision mode and a progressive lens is
placed on the nose piece, the symbol “P” is displayed at upper left corner
of the screen.
MODE
CYL
STEP
R
SPH CYL AXIS ADD
When the AUTO PROG setting is enabled (see Section 3.4, Setup), the
MULTI-FOCAL mode will be automatically activated. See Section 3.2.2
a), Measuring progressive addition lenses.
Use the center of the lens to detect the progressive
channel. The progressive channel may not be detected in
peripheral areas.
A progressive lens with a low ADD power may not be
recongized as a progressive lens.