4.2 ROUTINE MAINTENANCE
Do not touch the optical elements, such as the cover glass, with your
fingers. Keep them clean. Accuracy of the reading may be adversely
affected by dust or dirt.
Keep the dust cover over the instrument when the instrument is not in
immediate use.
Use a dry cloth to clean the cover, screen and the front panel. Use
diluted non-organic detergents for any heavy stains. Do not use organic
solvents, such as thinners, which may cause damage to the surface of
the instrument.
Disconnect the power cord if you do not plan to use the instrument for a
long time.
4.2.1 CLEANING COVER GLASS
Clean the cover glass periodically.
1) Remove the nose piece.
2) Clean the cover glass as follows:
Remove the dust on the cover glass using a blower (i.e., compressed
air). If the dirt still remains, gently wipe it off using a soft cloth or lens
cleaning accessories.
4.2 ROUTINE MAINTENANCE
Scratches on the cover glass may result in inaccurate
readings.