ー46ー
8-2.
Preventive maintenance to be carried out every week
WARNING
Stop operation when corrosion, damage, or deformation is found on
the chamber, chamber lid, hinges, hinge shafts, cover rings, lid hooks,
or lid hook catchs.
Contravention is dangerous with likelihood of death or serious accident due to
blowout during rise of the pressure. When an abnormality is found, contact the
dealer or business institution of our company.
Stop operation when damage such as cracks is found on the chamber
lid gasket.
Contravention is dangerous with likelihood of scalds due to steam leakage.
When an abnormality is found, contact the dealer or business institution of our
company.
Unplug and check the power plug regularly.
Failure to do so may cause a fire or burnout.
1. Cleaning the inside of the chamber and the water level sensor
Remove the grating carefully.
Thoroughly clean the inside of the chamber and the water level sensor using mild detergent
and cloth or other suitable material, then wash out the detergent by means of tap and drain
water.
Be careful not to distort the sensors in the chamber during cleaning.
<
Cleaning
>
For ensuring safe operation of the autoclave, carry out cleaning and inspection described below
every week as a guide.
2. Cleaning the lid gasket
Wipe off the surface of the chamber lid gasket using a damp soft clean cloth etc. without
applying force.
Be careful not to damage the lid gasket, while cleaning.
Wipe off opening of the chamber using a soft clean cloth etc.
3. Cleaning the outer covering
Remove contaminants using a soft cloth moistened with mild detergent.
Wipe off detergent using a tightly wrung-out damp cloth.