Guidance and manufacturer's declaration - electromagnetic immunity
The ACP-8 is intended for use in the electromagnetic environment specified below.
The customer or the user of the ACP-8 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment
- guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, con-
crete or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV
differential mode
± 2 kV
common mode
± 1 kV
differential mode
± 2 kV
common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
Voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
t
(>95% dip in U
t
)
for 0.5 cycle
40% U
t
(60% dip in U
t
)
for 5 cycles
70% U
t
(30% dip in U
t
)
for 25 cycles
<5% U
t
(>95% dip in U
t
)
for 5 sec
<5% U
t
(>95% dip in U
t
)
for 0.5 cycle
40% U
t
(60% dip in U
t
)
for 5 cycles
70% U
t
(30% dip in U
t
)
for 25 cycles
<5% U
t
(>95% dip in U
t
)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
or the ACP-8 requires continued
operation during power mains
interruptions, it is recommended
that the ACP-8 be powered from
an uninterruptible power supply
or battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields
should be at levels characteris-
tic of a typical location in a typical
commercial or hospital environ-
ment.
NOTE U
t
is the a.c. mains voltage prior to application of the test level.