25
Resetting FC-200
If there is no response to the pressing of a key or tapping of the display, you need to reset the FC-
200. There are 3 methods to reset the FC-200.
Soft Reset
With the soft reset, all running applications will be shutdown, and data you are editing will be lost;
however, installed programs and stored data are protected.
How to soft reset
1 Remove all cables or cards (CFs, SDs) that are currently mounted.
2 While pressing the [Alt] key, press the curser key to right and up direction for at least 2
seconds.
Hard Reset
If the instrument does not function properly after soft reset has been done, perform a hard reset.
A hardware reset will revert all settings to defaults and erase all RAM data.
How to hard reset
1 Remove all cables or cards (CFs, SDs) that are currently mounted.
2 Lightly press the Hard reset switch in the middle of the card slot with stylus pen for at least 2
seconds.
A hardware reset will erase all RAM data and applications.
Regular backups will ensure RAM data and applications can be restored.