8. Good Laboratory Practices (GLP)
The WP-81 keeps a record of the date and time of the last Conductivity,
pH and Temperature calibrations as part of GLP guidelines.
8.1 To recall GLP information on the display
1. Switch the meter on.
2. Select the GLP menu. ( F4:Setup F3:GLP)
3. Select F1:Recall from the menu.
4. The instrument model, firmware version number, and instrument serial
number are displayed, along with a prompt describing how to scroll
through the GLP information.
5. Press the key to sequentially scroll through the GLP information
for all parameters. Press the key to scroll back to previous data.
The sequence of information displayed is shown below. Press to
abort at any time.
WP81 V6.0 S1234
:More :Back
Cond Zero 0.00uS
@ 31/12/03 11:00
k=1.00 @ 2.76mS
@ 31/12/03 11:10
pH Asy 0.10pH
@ 31/12/03 11:40
pH Slope 100.0%
@ 31/12/03 11:50
Temp Offset 0.1
O
c
@ 31/12/03 12:00