F7
Overload Limit
Selects the desired formula which determines the point at which the
indicator shows overload. All selections are based on the primary unit
selected in F8.
"FS" = Full scale in primary units.
FS
FS + 2%
FS + 1d
FS + 9d
F8
Calib. Unit
Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
"1" = primary unit is lb. "2" = primary unit is in kg.
1
2
F9
Display
Divisions
Determines the desired weight increments. Value should be consis-
tent with legal requirements.
1
2
5
F10
Decimal Pt.
Determines location of the decimal point.
0 0.0
0.00 0.000
0.0000 00
F16
Zero Calibra-
tion
Places indicator into the zero calibration routine. Scrolling down with
the UNIT/SET key one level begins the procedure.
Press UNIT/SET
key to begin se-
quence
F17
Span Calibra-
tion
Places indicator into the span calibration routine. Scrolling down with
the UNIT/SET key one level begins the procedure. NOTE – press and
hold the UNIT/SET for three seconds for effect.
Press UNIT/SET
key to begin se-
quence
F18
View Calibra-
tion
Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed.
Scrolling down with the UNIT/SET key one level begins the proce-
dure.
Press UNIT/SET
key to begin se-
quence
F19
Key-in Zero
Allows you to key-in known zero calibration value in case of memory
loss in the field. Scrolling down with the UNIT/SET key one level be-
gins the procedure.
Press UNIT/SET
key to begin se-
quence
F20
Key-in Span
Allows you to key-in a known span calibration value in case of mem-
ory loss in the field. Scrolling down with the UNIT/SET key one level
begins the procedure.
Press UNIT/SET
key to begin se-
quence
F21
Factory Reset
This sub-menu will reset all parameters in the “F” and “A” menu to the
default settings. USE WITH CAUTION!
Press the
UNIT/SET key
twice to execute.
F24
Piece Count
Enable
Used to enable or disable the piece count feature. This feature must
be disabled for commercial applications.
"0" = Disable piece count "1" = Enable piece count
0
1
Page 4-2
Ne
p
tune 4500

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Transcell Technology NEPTUNE 4500 and is the answer not in the manual?

Transcell Technology NEPTUNE 4500 Specifications

General IconGeneral
BrandTranscell Technology
ModelNEPTUNE 4500
CategoryTouch Panel
LanguageEnglish

Related product manuals