CODE/NAME
DESCRIPTION
SELECTION LIST
F4
Zero Range
Selects the range (expressed as a percentage of full scale capacity)
within which the scale may be ze
roed. Note that the indi
cator must be
in standstill to zero the scale.
100%
1.9%
F5
Motion Band
Sets the level at which motion is detected. If motion is not detected,
the scale can process a Print or Zero command. Maximum value
varies depending on local regulations. Expressed as scale divisions
per second (d/s).
1d
3d
5d
10d
F6
Digital Filter
Averages weight readings to produce higher stability. Choose the
speed that works best for your application.
“FAST” = Fast “nnEd” = Medium “SLo” = Slow
FAST
nnEd
SLo
F7
Overload Limit
Selects the desired formula which determines the point at which the
indicator shows overload. All selections are based on the primary unit
selected in F8.
"FS" = Full scale capacity.
FS
FS + 2%
FS + 1d
FS + 9d
F8
Calib. Uni
t
Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
"1" = primary unit is lb. "2" = primary unit is in kg.
1
2
F9
Display
Divisions
Determines the desired weight increments. Value should be
consi
s
tent with legal requirements.
1
2
5
F10
Decimal Pt.
Determines location of the decimal point.
0
0.0
0.00 0.000
0.0000 00
F16
Zero
Calibr
a
tion
Places indicator into the zero calibration routine. Scrolling down with
the ZERO key one level begins the procedure.
Press
ZERO
key
to begin sequence
F17
Span
Calibr
a
tio
n
Places indicator into the span calibration routine. Scrolling down with
the ZERO key one level begins the procedure.
Press
ZERO
key
to b
egin sequence
F18
View
Calibr
a
tion
Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed.
Scrolling down with the ZERO key one level begins the pro
cedure.
Multi
-
point cal
Press
ZERO
key
to begin sequence
F19
Key
-
in Zero
Allows you to key-in known zero calibration value in case of memory
loss in the field. Scrolling down with the ZERO key one level begins
the
procedure.
Press
ZERO
key
to begin sequence
F20
Key
-
in Span
Allows you to key-in a known span calibration value in case of
mem
ory loss in the field. Scrolling down with the ZERO key one level
be
gins the procedure.
Press
ZERO
key
to begin sequence
F21
Fac
tory Reset
This sub
-
menu will reset all parameters in the “F” and “A” menu to the
default settings. USE WITH CAUTION!
Press the
ZERO
key twice to
ex
e
cute