13
If the calibration was not successful, one of the following error messages will appear.
"Err0" - The calibration test load or the keyed-in load is larger than the full capacity of
the indicator. Change the calibration test load or check the input data.
"Err1" - The calibration test load or the keyed-in load is smaller than 1% of the full ca-
pacity of the indicator. Change the calibration test load or check the input data.
"Err2" There is not enough signal from the scale platform to complete the calibration
process. Most common causes include incorrect weight sensor wiring, a mechanical
obstruction or a faulty (damaged) load cell.
Take the indicated action to correct the problem, and then perform a new calibration.
Exit calibration mode by moving the slide switch back to its original position. The display will go
through a digit check, and then settle into Normal Operating mode. All front panel keys will now return
to their normal mode of operation.