4
The keypad comprises five (5) function keys.
Marking
Keypad Function
Units
Selects the displayed unit of measure, i.e., lb or kg. This key can be disabled.
Zero
Zeroes the weight display reading, provided certain conditions are met
1
.
Net/Gross
Selects the weighing mode, i.e., Gross weight or Net Weight.
Tare
Establishes a system Tare, provided certain conditions are met
2
.
Print
Sends the displayed weight reading to the serial communication port, provided
certain conditions are met
3
.
1
Scale mustn’t be in motion or displaying an error message, e.g., overload. Scale must be in Gross mode.
Operation may be restricted by the Zero Reset Range setting (see F4 menu parameter).
2
Scale mustn’t be in motion or displaying an error message, e.g., overload. Tare weight must be greater
than Gross Zero.
3
Scale mustn’t be in motion or displaying an error message, e.g., overload.