5.2 Measurement Properties
5.2.1 Inuences on the Measurement
The pH measurement depends on the following parameters:
High temperatures of the measuring medium accelerate the ageing of the electrode.
Note the immersion depth: a too shallow immersion depth can lead to uctuating measured values.
Air bubbles on the sensor can cause measurement errors.
Alkali error: at pH values above pH12, alkali ions (Li+, Na+) can cause lower pH values to be displayed, as the
alkaliions are detected in addition to the H+ ions.
Flow can inuence the measured value.
5.2.2 Temperature Compensation
Temperature compensation is performed automatically by the integrated temperature sensor (Pt1000).