5. Meter comes
on, displays
software
version, and
goes to Meter
Set Up. Check
settings (see
Meter Set Up).
If meter does not turn on, open Battery Door and
check that the battery was inserted “+” with the
side facing up. Close Battery Door and repeat
Step 5. If meter still does not turn on, call for
assistance.
Caution!
• Batteries may explode if mishandled. Do not
dispose of battery in fire. Do not take apart
or attempt to recharge battery. Dispose
according to local regulations.
4. Close Battery
Door.
Blood Glucose Monitoring System Specifications
Result Range: 20-600 mg/dL
Sample: 0.4 microliter (0.4 µL) Fresh capillary whole blood
Test Time: Results in 4 seconds
Result Value: Plasma equivalent values
Assay Method: Electrochemical
Power Supply: One 3V lithium battery #CR2032 (non-rechargeable)
Battery Life: Approximately 1,000 tests or 3 years
Automatic shut-off: After two minutes of non-use
Weight: 1.5 ounces
Size: 2.0” x 3.5” x 0.75”
Memory Size: 400 results
Operating Range (Meter & Test Strips): Relative Humidity: 10-93% (Non-condensing)
Temperature: 57°F - 104°F
Hematocrit: 25-60%
~ Use within specified environmental conditions only.
Chemical Composition
TRUE FOCUS Test Strips: Glucose dehydrogenase-FAD (Aspergillus sp.), mediators, buffers and stabilizers.
TRUE FOCUS Control Solution: Contents: D-glucose, preservative, red dye and viscosity-adjusting agent.