EasyManua.ls Logo

Trividia TRUE FOCUS - Testing Blood; Obtaining a Blood Sample

Trividia TRUE FOCUS
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22 23
TESTING BLOOD
Obtaining a Blood Sample
Refer to lancing devices Instructions for Use for more detailed instructions on using the lancing device.
Important Notes Regarding Forearm Testing
6
~ Alternate site testing results cannot be used for continuous glucose meter calibration or for insulin dose
calculations.
~ Check with your doctor or healthcare professional to see if forearm testing is right for you.
~ Results from forearm are not always the same as results from fingertip.
~ Some lancing devices include a special end cap for alternate site (forearm) testing. Check lancing device
Instructions for Use.
~ Use fingertip instead of forearm for more accurate results:
Within 2 hours of eating, exercise, or taking insulin,
If your blood sugar may be rising or falling rapidly,
If your routine results are often fluctuating,
If you are ill or are under stress,
If your forearm results do not match how you feel,
If your blood sugar may be low or high,
If you do not notice symptoms when blood sugar is low or high.
Caution!
Wash your hands thoroughly with soap and warm water before and after handling the meter, lancing
device, lancets or test strips.
If the meter is being operated by a second person who gives testing assistance, the meter and the lancing
device should be cleaned and disinfected before use by another person.
For instructions on how to clean and disinfect the meter and lancing device, see Meter Cleaning and
Disinfecting and Lancing Device Cleaning and Disinfecting in this Owner’s Booklet.
ALL parts of your blood glucose monitoring system could carry blood-borne pathogens after use, even after
cleaning and disinfecting.
2,3
Cleaning and disinfecting the meter and lancing device destroys most, but not
necessarily all, blood-borne pathogens.
DO NOT share your meter or lancing device with anyone, including family members.
Do not reuse lancets. Reuse of devices labeled for single-use may result in product contamination and
patient infection.
Used test strips and lancets are considered biohazardous. Dispose used test strips and lancets into
appropriate waste container.
Do not use venous blood for testing.
To help prevent false high results, wash hands before using the blood blucose monitoring system to test
blood, especially after fruit has been handled.
Tfocus_OB_GutsENG_UE4TVH03r43.indd 22-23 4/19/18 07:42