EasyManua.ls Logo

Troxler 3411-B - Page 39

Troxler 3411-B
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 3411-B Advanced Control Unit 3–3
3. ADVANCED OPERATION
DENSITY OFFSET
To access the Special functions, press SPECIAL.
Press the down arrow key once to access the Offset function. Press
START/ENTER to display:
Offset: Density
( or ENTER)
Press START/ENTER.
Dens. Offset OFF
Want to Enable?
To enable the Density Offset function, press ON/YES.
D off= 0.0 PCF
( or ENTER)
Input the difference between the gauge and actual density readings.
To input a minus sign (for a negative offset), press the down arrow
first. To scroll through the numerals, press up and down arrows.
To select the next digit and/or exit, press START/ENTER.
The display will be:
Dens. Offset ON

Other manuals for Troxler 3411-B