Model 3411-B Advanced Control Unit 4–9
4. SPECIAL FUNCTIONS
CALIBRATION CONSTANTS
The Calibration Constants function allows the operator to change
the mathematical constants used for calculating a test result. If the
gauge has been repaired or the memory has been lost, the constants
must be verified or re-entered.
NOTE
Each gauge contains a unique set of constants. The
constants used in one gauge will not work in another
gauge! The constants for your gauge are developed
when the gauge is calibrated and are reflected on the
most recent calibration sheet.
For more information on the Calibration Constants feature, refer to
page 1–4.
MEMORY RESET
NOTE
This function is for authorized service personnel only!
CAUTION
Memory Reset erases all data stored in the gauge and
sets all constants, except calibration constants, to the
default values.
TEST READING
NOTE
This function is for authorized service personnel only!