Model 3411-B Advanced Control Unit 4–3
4. SPECIAL FUNCTIONS
To perform a stat test, place the gauge on the reference standard
block in the standard count position (see page 2–8).
To access the Special functions, press 〈SPECIAL〉.
To access the Stat Test feature, press the down arrow twice and
press 〈START/ENTER〉.
press START for
20 m. Stat Test
To begin the twenty counts, press 〈START/ENTER〉.
The gauge will display the stat test count progress as shown below.
- STAT TEST -
Rdg.#:xx xx sec
Upon completion of the stat test, the gauge displays the pass/fail
status. If the stat test fails, repeat the test twice more. If two out of
three stat tests fail, contact Troxler Technical Support (see page
B-11). If the stat test passes, the display is:
D: PASS M: PASS
↑↓ to view data
To view the stat test data, use the up and down arrow keys.
Dens. R = x.xxx
↑↓ to view data
Dens. Avg. x
↑↓ to view data