Specifications 45
Power Requirements:
Four AA-size batteries (included) or
AC Adapter p/n 801761
Input: 90 to 240 VAC, 50 to 60 Hz
Output: 9 VDC, 2A
1
Temperature compensated over an air temperature range of 40 to 150°F (5 to
65°C).
2
The accuracy statement of ±3.0% of reading or ±3 ft/min (±0.015 m/s), whichever is
greater, begins at 30 ft/min through 9999 ft/min (0.15 m/s through 50 m/s).
3
Pressure velocity measurements are not recommended below 1000 ft/min (5 m/s)
and are best suited to velocities over 2000 ft/min (10.00 m/s). Range can vary
depending on barometric pressure.
4
Accuracy is a function of converting pressure to velocity. Conversion accuracy
improves when actual pressure values increase.
5
Accuracy with instrument case at 77°F (25°C), add uncertainty of 0.05°F/°F
(0.03°C/°C) for change in instrument temperature.
6
Accuracy with probe at 77°F (25°C). Add uncertainty of 0.1% RH/°F (0.2% RH/°C)
for change in probe temperature. Includes 1% hysteresis.
7
Overpressure range = 190 in. H
2
O (7 psi, 360 mmHg, 48 kPa).