EasyManua.ls Logo

TSI Instruments SMPS 3938 - Electrostatic Classifier User Calibration Screen-Sheath Flow Tab

TSI Instruments SMPS 3938
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Operation 5-43
2. Disconnect the Sheath Flow + tubing from the DMA and connect it to
a reference flowmeter with a flow accuracy of ± 2% or better over the
range of 2 to 30 L/min, such as a Gilibrator
®
meter.
3. Connect the Sheath Flow port on the classifier to the Sheath Flow
port on the DMA if not already connected.
4. From the classifier Setup screen navigate to Calibration > User
Calibration. Select the Sheath Flow tab.
Figure 5-28
Electrostatic Classifier User Calibration ScreenSheath Flow Tab
5. To check calibration at a specific flow rate, set the sheath flow to the
flow rate of interest by pressing the corresponding Setpoint button. The
Raw and Classifier data fields populate, the Reference cell for that
row is activated, and an onscreen keypad appears.

Table of Contents

Other manuals for TSI Instruments SMPS 3938

Related product manuals