Maintenance, Service, and Troubleshooting 6-11
Replacing the Flowmeter
The flowmeter could be damaged if flow was run in the wrong direction for
a significant period of time or if the instrument was dropped or otherwise
mishandled. A “Sheath flow error message on the display screen may
indicate a damaged flowmeter. To replace the flowmeter, follow these
instructions:
1. Power off the classifier.
2. Remove the cover following the instructions at the beginning of this
section (see Removing the Cover).
3. Using a 9/64-inch hex wrench, remove the three 8-32 x 3/8-inch
screws holding the upper flow manifold in place. Save the screws for
reuse.
Note: Use the clearance holes in the upper flange of the instrument
case to access these screws.
4. Lift the upper flow manifold vertically until it separates from the
flowmeter (black).
5. Pull the flowmeter upwards until it separates from the main flow
manifold.

Table of Contents

Other manuals for TSI Instruments SMPS 3938

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments SMPS 3938 and is the answer not in the manual?

TSI Instruments SMPS 3938 Specifications

General IconGeneral
BrandTSI Instruments
ModelSMPS 3938
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals