EasyManua.ls Logo

TSI Instruments SMPS 3938 - Page 213

TSI Instruments SMPS 3938
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Computer Interface C-7
Table C-2
Electrostatic Classifier Firmware Commands List (select)
Command
Title
Description
RMERRORS
Read Classifier
Errors
RMERRORS returns the classifier errors.
Example: RMERRORS returns a 32 bit integer. If the bit is set, there is
an error.
Bits 0
Impactor delta pressure > 5cm H
2
O but no impactor present
1
N/A
2
Impactor delta pressure < 5cm H
2
O but impactor is present
3
Impactor dP > 9.75 kPa
4
No DMA detected
5
Sheath flow rate 2% greater than setpoint for longer than
30 seconds
6
High voltage outside range of setpoint
7
Built in self-test failed
8
Case temp > 50 °C or < 5 °C
9
X-ray and radioactive neutralizer both installed
10
X-ray runtime hours too high
11
Corrupt detector data
12
N/A
13
N/A
14
N/A
15
N/A
16
aP sensor < 17 kPa and > 128 kPa
17
N/A
18
Blower over current
19
Detector efficiency file missing or not monotonic
20
Charge correction file missing
21
Data logging error
22
N/A
23
Not used
24
Detector firmware not supported
25
Sheath flow rate is stabilizing from a setpoint change

Table of Contents

Other manuals for TSI Instruments SMPS 3938

Related product manuals