Range Resolution Accuracy Overload Protection
-328˚F to 999˚F
(-200˚ to 999˚C)
0.1°F (0.1˚C)
±(1.0% +3.6˚F)
±(1.0% + 2.0˚C)
30V RMS
1000˚F to 2462˚F
(1000˚ to 1350˚C)
1°F (1˚C)
• Disconnect test lead probes from voltage source and meter.
• Move Input Jack Locks to “TEMP” setting.
• Use K-Type thermocouple temperature probes only.
• Stated accuracy does not account for thermocouple accuracy.
Features:
HOLD
MAX/
MIN
PEAK
Default = T1
Press x1 = T2
Press x2 = T1 - T2
Press and hold =
˚C
Temperature F˚/C˚
DL429B
600V 600A
CAT Ill
CATIII
600V
MAX
30V
T2
T1
TEMP
INRUSH
BT
T
1
-T
2
NCV
MAX/
MIN
PEAK
DC
Zero
INRUSH
T
e
m
p
/
°
F
°
C
μ
A
A
V
H
z
/
D
u
t
y
MFD
RANGE
True RMS
LINK
T
e
m
p
/
°
F
°
C
TEMP
• Buzzer sounds at less than < 40Ω.
WARNING
• Do not measure resistance on a live circuit.
Features:
MAX/
MIN
PEAK
HOLD
Open Circuit V <1.00V Overload Protection
Threshold Approx. <40Ω 600V RMS
Continuity
Default =
Press x1 =
Press x2 =
Press x3 =
DL429B
600V 600A
CAT Ill
CATIII
600V
MAX
30V
T2
T1
TEMP
INRUSH
BT
T
1
-T
2
NCV
MAX/
MIN
PEAK
DC
Zero
INRUSH
T
e
m
p
/
°
F
°
C
μ
A
A
V
H
z
/
D
u
t
y
MFD
RANGE
True RMS
LINK

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the UEi DL429B and is the answer not in the manual?

UEi DL429B Specifications

General IconGeneral
BrandUEi
ModelDL429B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals