Questions and Answers

  • B
    Brittany BrooksSep 8, 2025
    Why doesn't HV ON work even after pressing the ON button on my Ulvac Power Supply?
    • W
      William BriggsSep 8, 2025
      If HV ON does not work even after pressing the ON button on your Ulvac Power Supply, it could be due to several reasons: * HV SWEEP not working: Consult Section 5. * INTERLOCKS connector issue: Check the INTERLOCKS connector, specifically the contact between Pin 7 and Pin 8, referring to Fig. 4.7 for wiring and contacts. * KEY switch in LOCK state: Ensure the KEY switch is in the LOCAL mode, not LOCK mode.
  • B
    Brandon ColemanAug 23, 2025
    How to troubleshoot an Ulvac HPS-1600F that will not switch ON?
    • D
      dorismcgrathAug 23, 2025
      If the Ulvac Power Supply cannot be switched to ON, consider the following possible causes and solutions: 1. **No power received**: Check the power supply in your fuse box and ensure it is receiving power. 2. **Panel interlock active**: Verify that the panel is properly in place and that the panel mounting screws are not loose. 3. **MAIN circuit breaker broken**: Determine if the breaker trips even when the power supply in the fuse box is in the OFF state. If so, replace the breaker.
  • J
    Joseph KrauseSep 5, 2025
    What causes the Gun controller indicator to turn off with the key switch at LOCAL on my Ulvac Power Supply?
    • M
      Megan WellsSep 5, 2025
      If the Gun controller indicator is turned off with the key switch at LOCAL on your Ulvac Power Supply, check the connections of the interlock on the equipment. Also, check whether there is a ground fault of the interlock signal.
  • A
    Alexander GreerSep 3, 2025
    Why does the HV OFF lamp not light up on my Ulvac HPS-1600F?
    • D
      Danielle JonesSep 4, 2025
      If the HV OFF lamp does not light up on your Ulvac Power Supply, check the following: * Ensure that the INTERLOCK is released. Check each interlock indicator of the Gun controller and satisfy the interlock conditions if any interlock is off. * Confirm that the Gun controller key switch is not in the LOCK position. If it is, set it to the LOCAL position.
  • A
    Amanda GordonSep 2, 2025
    What to do if the Gun controller panel does not light up on Ulvac Power Supply?
    • S
      Sheila WhiteSep 2, 2025
      If the Gun controller panel does not light up on your Ulvac Power Supply, check the connection of the POWER and CONTROL cables.
  • G
    greynoldsSep 1, 2025
    Why can't the acceleration voltage be applied as set on my Ulvac Power Supply?
    • J
      jjonesSep 1, 2025
      If the acceleration voltage cannot be applied as being set on your Ulvac Power Supply, consider the following: * Check if HV SWEEP is enabled. If so, refer to Fig. 4.6 and set HV SWEEP again. * Verify that you are not attempting to set a high voltage on the side of source 2, as it is not possible.
  • D
    David SuttonAug 30, 2025
    Why is remote operation not available on my Ulvac Power Supply?
    • A
      Alexander JosephAug 30, 2025
      If the remote operation is not available on your Ulvac Power Supply, check the following: * Ensure it is not in LOCAL operation mode by checking the LOCAL/REMOTE settings. * Verify that an external signal is being input by checking the program connector and the external signal sources.
  • D
    David SnyderAug 29, 2025
    Why does the fuse blow when the high voltage is switched ON in my Ulvac HPS-1600F Power Supply?
    • D
      deckerangelaAug 29, 2025
      If the high voltage is switched to ON and the fuse is blown in your Ulvac Power Supply, it could be because the beam current command signal is maximized. Check the command signal of the beam current. A large inrush current may pass at the instant when the high voltage is switched to ON if the signal is maximized.
  • P
    Pamela Snow PhDAug 27, 2025
    Why does no beam current pass through my Ulvac HPS-1600F?
    • R
      Ryan GrahamAug 28, 2025
      If no beam current passes with your Ulvac Power Supply, verify the following: * Check that a high voltage is applied. * Confirm that it is not in DEGAS mode and release it if it is. * Check the wiring of the filament line. Also, consider potential issues on the loading side, such as poor insulation or contact of the filament line due to assembly failure of the EB gun, or dirt on the isolation insulator.
  • J
    Jacqueline Jones MDAug 26, 2025
    Why is the Gun controller READY indicator not lighting up on my Ulvac HPS-1600F?
    • S
      sherry94Aug 27, 2025
      If the Gun controller READY indicator on your Ulvac Power Supply does not light up, check the following: * **SERVICE KEY settings**: Ensure the SERVICE KEY is in the ENABLE state, not LOCK mode. * **Fuse indicator**: Check the fuse indicator of the power supply main unit. * **Fan rotation**: Make sure the fan is rotating and remove any foreign substances that may be stopping it.