3.5 Trigger Hold-off
Trigger hold-off can observe the complex waveforms (such as pulse train). Hold-off time is the amount of time
the oscilloscope waits before re-enabling the trigger circuit. During the hold-off period, the oscilloscope will not
trigger until the hold-off time is over. For example, a set of pulse train, which is required to trigger on the first
pulse, the hold-off time can be set to the pulse train width.
Press the HORI MENU on the horizontal control area and then adjust the Multipurpose knob to set the trigger
hold-off time. Input a combined waveform to CH1, and adjust the trigger hold-off time until the waveform can be
triggered steadily, as shown below:
Chapter 4 Sampling System Settings
Sampling is taking analog input signal and converts it into discrete points by using the analog to digital converter (ADC).
Press the ACQUIRE key to enter the sample menu.
Descriptions
Sampling in the normal way
Sampling in peak detection mode
Sampling in high-resolution mode
Sampling in envelop mode
Sampling in an average way
In the average acquisition mode, you can adjust the Multipurpose
knob to set the average number of times, which can be set to 2n,
and n is an integer from 1 to 13.
Set the memory depth to automatic, which is the normal memory depth
Set storage depth to 28kpts
Set storage depth to 280kpts
Set storage depth to 2.8Mpts
Set storage depth to 28Mpts
Set storage depth to 56Mpts
Options
Normal
Peak
High resolution
Envelope
Average
2~8192
Auto
28k
280k
2.8M
28M
56M
Functions
Sampling Mode
Average
Storage Depth
19