Accuracy Specifications (2)
C. Continuity, Diodes & Resistance Test
Range Resolution Accuracy Overload
Protection
Continuity Test
(400.0 )
0.1
Approximate
100
Diode 1mV N/A
400
0.1
(1.2%+2)
4k 1
40k 10
400k 100
(1%+2)
4M
1k
(1.2%+2)
40M
10k
(1.5%+2)
600Vp
Remarks:
l Continuity Test (400.0 ) Range:
Buzzer beeps continuously.
Open circuit voltage approximate 0.45V.
l Diode Range:
Open circuit voltage approximate 1.48V.
Displays approximate forward voltage drop
reading 0.5V~0.8V.
l 400 ~ 40M Range:
Open circuit voltage approximate 0.45V.