Remarks:
At 5A range: Continuous measurement is allowed.
At >5A range : For continuous measurement 10 seconds and interval not less than 15 minutes.
Displays:
True RMS (applicable to the range of 10%~95%)
AC peak factor 3.0.
Input short circuit has around less than 30 remaining digits, it will not affect accuracy.
AC+DC measurement accuracy = range's accuracy + 1%
42
Fuse 500mA, 250V, fast
type, 5x20mm.
Fuse 10A,250V,
fast type, 5x20mm.
Model UT803: OPERATING MANUAL
D. AC Current
l
l
l
Range Resolution Accuracy Overload Protection
600µA
6000µA
60mA
600mA
10A
0.1µA
1µA
0.01mA
0.1mA
10mA
40Hz~10kHz: (1.0%+5)
>10kHz~15kHz: (2%+5)
40Hz~10kHz: (1%+5)
>10kHz~15kHz: (3%+5)
40Hz~5kHz: (2.0%+6)