EasyManua.ls Logo

UNI-T UTD1000 Series - Page 46

UNI-T UTD1000 Series
101 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...

UTD1000 User Manual
Example 1 :
To display the measurement peak-to-peak value of
Channel A, follow the steps below :
1.Press [A] to turn on Channel A.
2.Press [SCOPE] to enter the frequently used function
menu.
3.Press [F2] to enter the parameter measurement
options.
4.Press [F1] to select customized parameter for
parameter measurement.
5.Press [F2] to select parameter 1 as the customized
parameter.
6.Press [F3] then turn the jog dial to select peak-to-
peak value, then press the jog dial once to confirm. The
screen will exit from the current menu.
7.Press [F4] to select Channel A.
8.Press [F5] to set indicator to parameter 1.
The peak-to-peak value is displayed in the top right
corner, as shown below :
Figure 2-22 Selected parameter measurement
Note : Customized parameters are for quick
parameter measurement. This device has 27
measurement parameters. In general use, the
user is not required to measure all parameters.
Only a few have to be tested. In this case, set the
parameters you require as customized parameters.
All such parameters will be displayed on the screen.

Related product manuals