UTD2000 Series User Manual
32
Notice:
(1) Please select real-time sampling mode when observing single signal.
(2) Please select equivalent sampling mode when observing high frequency periodic
signal.
(3) Please select peak detection mode when hoping to observe the signal envelope for
any modulation signal.
(4) Please select average sampling mode when hoping to reduce the random noise in
displayed signals.
Explanation of nouns
Real-time sampling mode:
Data sampling at the real-time sampling rate of the system. This is used to observe
any waveform within the sampling rate of the system.
Equivalent sampling mode:
Data sampling at higher than maximum sampling rate of the system. This is used to
observe any periodical waveform at beyond the maximum sampling rate.