UTD2000 Series User Manual 
55 
 
(2) Connect the signal to ensure stable display of waveform.   
(3) Improving trigger by setting trigger coupling. 
①  Press TRIG MENU in the trigger zone to display the trigger setup menu. 
②  Set trigger coupling to Low Frequency Holdoff or High Frequency Holdoff.  Low 
frequency  holdoff  is  seting  up  a  high-pass  filter.  It  filters  low  frequency  signal 
component under 80kHz and allows high frequency signal component to pass through. 
High frequency holdoff is setting up a low-pass filter. It filters high frequency signal 
component over 80kHz and allows low frequency signal component to pass through. 
By setting Low Frequency Holdoff or High Frequency Holdoff,  you can hold off low 
frequency or high frequency noise respectively and achieve a stable trigger. 
(4) Reducing display noise by setting the sampling mode. 
①  If the signal being measured is stacked with random noise and the waveform is too 
coarse as a result, you can use the average sampling mode to eliminate random noise 
display and reduce the size of waveform for easy observation and measurement. After 
getting the mean, random noise is reduced and details of the signal are clearer. Follow 
the steps below: 
Press ACQUIRE in the menu zone of the front panel to  display the sampling setup 
menu. Press F1 to set acquisition mode to AVERAGE, then press F1 to adjust the 
average number of  times in multiples of 2, i.e. 2 to 256, until you get the  desired 
waveform display that meets observation and measurement requirements. (See the 
figure below) 
 
 
Figure 12-5 Signal noise held off 
 
Note: In average sampling mode, the waveform display will update at a slower rate. 
This is normal. 
 
 
 
Example 5: Using the cursors for measurement 
Your  oscilloscope  can  measure  28  waveform  parameters  automatically.  All  auto 
parameters can be measured with the cursors. 
By using the cursors, you can quickly measure the time and voltage of a waveform.