Chapter 8: Trouble-shooting
Troublecode&
phenomenon
Possible Cause
Correction
Hardware or
software error
Turn off and on again.
If the error still exists,
please contact the local
distributor.
E0, E1
Meter used in
bright sunlig-
ht
Bright sunlight disturbs
optical system of meter.
Used it in room-light or
shade. Turn off and on
again. Repeat test with
new test strip.
E2
Temperature
out of Range
Allow meter to adjust to
temperature declared in
the Manual and wait un-
til it shake down. Turn
off and on again. Repe-
at test with new test
strip.