Chapter 9: Maintenance
1. Routine maintenance and cleaning
The cleaning must be done after appl-
ying superfluous blood sample every time.
Hold the bottom of meter with left hand and
lift holder, then the strip holder could be re-
moved and cleaned. Wipe away the rest
blood or dirt left on the holder or meter usi-
ng gauze with disinfectant (diluenting 10%
home bleacher, about 0.6% hypochlorite).
If there is superfluous blood on the op-
tical detector, clean it by cotton sticker with
little detergent. Wipe the strip holder and
optical detector completely by cotton cloth
without hemp. Avoid any liquid permeating
the meter anytime.
Do not polish the surface of optical det-
ector with any polished or corrosive deterg-
ent to avoid the damage of meter and optic-