when display shows Filt.1~Filt.7.
Select Filt.7 will get strongest filter effect.
- To disable Animal Weighing Function, press TARE
when display shows Filt.0
f. After TARE is pressed, indicator display F7
g. Press MODE to enter next set up or ON/ZERO to restart
the scale
4.2.5 Select Serial Output Mode
There are four selectable transmission modes to choose
according to end user preference. For which Conti is
continuous transmission when weight is stabilized,
Auto sends data once when weight is stabilized and
P_Out sends data in table form when press M+.
OFF disable Serial transmission
a. Scale is off
b. Press and hold TARE, then press ON/ZERO
c. Scale displays F1
d. Press TARE six times to F7
e. Press MODE to shift between Conti, Auto, P_Out or
OFF
- For continuous sending, press TARE when display
shows Conti
- For sending data once, press TARE when display