4.2.2 Select Auto Power Off Setting
The scale is equipped with Auto Power Off Function. The
scale will power off automatically if it is not used
for 4 minutes. Follow the below steps to enable/disable
Auto Power Off Function.
a. Scale is off
b. Press and hold TARE, then press ON/ZERO
c. Scale displays F.1
d. Press TARE three times
e. Scale displays F.4
f. Press MODE to shift between 0._OFF and 4._OFF
-To enable Auto Power Off Function, press TARE when
4._OFF appears
-To disable Auto Power Off Function, press TARE when
0._OFF appears
g. Press ON/ZERO to restart the scale or TARE for next
function setup
4.2.3 Select RS-232 Baud Rate & Protocol
This scale is equipped with 4 RS232C output baud rate
and 2 transmission protocol settings for user to select.
a. Scale is off
b. Press and hold TARE, then press ON/ZERO
c. Scale displays F.1
d. Press TARE four times
e. Scale displays F.5
f. Press MODE to select baud rate of 4800, 9600, 19200,
38400
g. Press and hold MODE for 2 seconds to enter
transmission protocol setup
h. Press MODE to select protocol of N81 or E71
i. When the preferred transmission protocol shows on
display, press ON/ZERO to confirm and restart the
scale or TARE for next function setup
NOTE:
The default data output format of the scale is
Baud Rate = 9600
Data Bit = 8
Stop Bit = 1
Parity = None
4.2.4 Select Data Transmission Mode
This scale provides 2 types of transmission mode
depending on the requirement and environment.
a. Scale is off
b. Press and hold TARE, then press ON/ZERO
c. Scale displays F.1