24
Replacing the circuit board
26.SetuptheBVCattheworkplace.
27.ConnecttheBVCtomains.
28.SwitchontheBVC.
29. Do not touch the touch panel, other‑
wise it takes several minutes until the
keys work correctly!
IMPORTANT!
> Adjusttheliquidlevelsensorafterreplacing
thecircuitboard.
> Checkthefunctionoftheliquidlevelsensor.
IMPORTANT!
> Checktheelectricalsafetyofthedevice
accordingtoIEC61010andnationalregula‑
tionsafterrepair.
Only BVC professional / G
30.AllLEDsofthetouchpanelarelit.
31.Waitapproximately1minuteuntilthe
pumpstarts.
32.Adjustsuctionpowerifnecessaryby
usingkey“+”or“‑”.