page 16 of 76
Type
BVC control / G
BVC professional / G
Pump
Maximum pumping speed
(ISO 21360) at 50/60 Hz
cfm
(m
3
/h)
0.4/ 0.5
(0.7 / 0.85)
Ultimate vacuum (absolute)
Torr
(mbar)
112
(150)
Maximum permissible inlet pressure
(absolute)
psi
(bar)
16
(1.1)
Maximum permissible outlet pressure
(absolute)
psi
(bar)
16
(1.1)
Rated power
hp
(kW)
0.054
(0.04)
No-load speed 50/60 Hz rpm 1500 / 1800
Control range
Torr
(mbar)
112 - 637
(150 - 850)
Permissible ambient temperature
storage / operation
°F
(°C)
14 to 140 / 50 to 104
(-10 to +60 / +10 to +40)
Maximum installation altitude m 2000 above mean sea level
Permissible relative atmospheric
moisture during operation
(no condensation)
% 30 to 85
Maximum permissible range of supply
voltage: Attention: Observe speci-
cations of rating plate!
100-120 V~ ±10% 50/60 Hz
230 V~ ±10% 50/60 Hz
Rated current at:
100-120 V~ 50/60 Hz
230 V~ 50/60 Hz
A
A
1.9 / 1.4
0.7 / 0.5
Fuse
2 x 2.5 AT, 0.19 x 0.78 in
(5 x 20 mm)
Motor protection thermal cutout
Degree of protection IEC 529 IP 40
A-weighted emission sound pressure
level* (uncertainty K
pA
: 3 dB(A))
dB(A) 45
Technical data
* Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 230V/50Hz and ulti-
mate vacuum with exhaust tube at outlet.