Do not modify the Duration, Settling, and Temperature
settings unless instructed to do so. The sensor element may be permanently
damaged from overheating.
CAUTION!
Type PUR and press ENTER to proceed. The maximum interval is 14400 minutes (10 days).
>pur
Interval Purge : OFF ?
Interval : 720 min ?
Power-up Purge : OFF ?
Duration : 240 s ?
Settling : 240 s ?
Temperature : 180 'C ?
>
To activate the new interval settings immediately, reset the probe.
When chemical startup purge is enabled, wait for 6 minutes after startup before
taking measurements. The output channels are locked for the first operation
minutes to the initial measured values.
4.7 Extra sensor heating (Xheat) commands
HMP155 extra sensor heating follows the factory default values. You can enable or disable the
function, change the RH limit, and define the heating temperature and duration of this
function.
The output channels are locked during extra sensor heating and cooling period.
For meteorological applications, the extra sensor heating function is rarely used
as radiation shield prevent water droplets from going directly to humidity sensor.
4.7.1 XHEAT command
You can enable or disable the extra sensor heating with the XHEAT command.
Chapter 4 – Operation
51

Table of Contents

Other manuals for Vaisala HUMICAP HMP155

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala HUMICAP HMP155 and is the answer not in the manual?

Vaisala HUMICAP HMP155 Specifications

General IconGeneral
Humidity Measurement Range0 to 100 %RH
Operating Humidity Range0 to 100 %RH
Protection ClassIP65
Measurement ParametersHumidity and Temperature
Temperature Measurement Range-40 to +60 °C
Temperature Accuracy±0.2 °C
Operating Temperature Range-40 to +60 °C
Humidity SensorVaisala HUMICAP®

Related product manuals