USER'S GUIDE_______________________________________________________________________
20 ___________________________________________________________________ M210194EN-A
6. To output the average pressure reading in the units of inHg .
>form <cr>
4.2P""UUU#r#n
?2.4P""UUUU #r #n <cr>
>unit inHg <cr>
P unit : inHg
>send <cr>
30.1234 inHg
>
7. Adding a text field, giving the address and outputting the
pressure reading from a identified barometer.
>form <cr>
4.2P""UUUU #r #n
? "Barometer " ADDR""4.2P""UUU#r#n<cr>
>addr 7 <cr>
Address : 7
>send 7 <cr>
Barometer 07 1020.30 hPa
>
8. To output the average pressure reading and the error status.
>form <cr>
4.2P""UUUU #r #n
?4.2P""UUU""ERR#r#n<cr>
>send <cr>
1020.30 hPa 0 (no error found)
>send <cr>
1020.30 hPa 1 (some error has occurred)
>
9. To output the transducer values (P1,P2,P3) separately and the
average value (P). Indicates errors if the maximum pressure
difference between the transducers is exceeding the defined value.
The two-or three-digit error field in the end of the line is expressing
the error state of the transducers. Number 1 means error and number 0
normal state. The first digit shows state of transducer 1 (P1), second
digit shows state of transducer 2 (P2) and the third digit shows
transducer 3 (P3).
>form <cr>
4.2P""UUUU #r #n
?4.2P1""P2""P""UUU""ERR#r#n<cr>
>send <cr>
1020.30 1020.32 1020.31 hPa 00
>send <cr>
1020.30 1022.30 1021.30 hPa 11 (Pd max error)
>