Vante ® Model 4160
20
2.12. General process information
NOTE: In hot, cold, or humid environments, work slowly and check the sealing head and
tubing frequently to ensure that they are clean and dry. Frequently verify the
quality of the tubing seals when working in hot, cold, or humid environments.
NOTE: Temperature and dwell time settings vary for different tube materials and sizes.
Adjust and validate your process settings for each application in your facility.
NOTE: The 4160 system displays a warning when the internal compressor runs for
5 minutes or longer. The warning states that the system is at risk of overheating.
Press OK to close the warning message. The compressor design uses a PTC
thermistor to automatically prevent the internal compressor from overheating.
WARNING: Do not place fingers or objects near or between the sealing head jaws when the
controller power is on. Severe burns can occur.
CAUTION: Keep the outside of the tubing and equipment clean and dry.
Invisible buffer paragraph
CAUTION: Allow the tubing to cool completely before releasing the sealing head. Releasing
the sealing head before the tubing cools can result in fluid leak.
NOTE: Vante provides system customization as needed. If your organization requires
additional controls in your tube sealing process, contact Vante for assistance.