Vante ® Model 4160
21
2.13. Troubleshooting
Table 2.1 4160 TPE sealing system troubleshooting guide
Problem
Possible Cause
Solution
Not sealing
No power to sealer
Make sure the unit is plugged in
and the power switch is turned
"ON".
Ensure the power cord is not
damaged.
Blown fuse
Replace the fuse. Refer to Fuse
Replacement on page 26.
Error message
Heater or sensor failure.
Send to Vante for repair.
Bad or incomplete seal
Incorrect settings
Verify the correct material is
selected.
Verify that the temperature, timing,
and cooling settings are appropriate
for the material and sealing
conditions.
Call Vante customer service for
more information.
Very slow sealing
cycle
Air supply is disconnected
Verify that the air supply tube is
connected to an appropriate source.
Uneven cooling
between heating plates
Possible obstruction in air
supply
Look for any kinks in the black air
hose leading to upper heating plate.
If the air supply is clear, contact
Vante customer service for more
instructions.