CAUTION
Isopropylalcoholisammableandmayeasilyignite
Caution! Hot surface!
Follow the safety instructions of the isopropyl alcohol manufacturer!
X Place all parts of the Cooling Unit
(except the Cap Lock) and the Filling
Chamber Screen in isopropyl alco-
hol for a maximum of 1/2 hour.
X If necessary, wipe the Cap Lock with
isopropyl alcohol.
X Rinse all parts under running water
for at least one minute.
X Let all parts dry completely.
6 CLEANING/HYGIENE
6.2.3 Check the Cooling Unit Parts and the Filling Chamber Screen
X Check all parts for damaged surfaces, cracks, softening, hardening, em-
brittlement, contamination or discoloration.
X Discard damaged parts.
6.2.4 Assembling the Cooling Unit
CAUTION
Plant particles in the Cooling Unit
Clogging of the Cooling Unit
Inhalation of plant particles
Insert the Screen correctly into the Cooling Unit Cap.
X Assemble the parts in reverse order.
X Insert the small Base Seal Ring completely into the Cooling Unit Cap.
Ö The small Base Seal Ring is no longer visible.
X When assembling, make sure that Seal Rings are properly seated.
X Insert Screen in the groove provided for this purpose.