6 CLEANING/HYGIENE
6.3 Filling Chamber
6.3.1 Cleaning the Filling Chamber
NOTICE
Material damage caused by improper cleaning
Penetration of isopropyl alcohol into the Vaporizer
Do not moisten cotton swabs too much.
NOTICE
Material damage caused by improper cleaning
Penetration of water into the Vaporizer, the USB Charger or the
USB-C Cable.
Do not immerse in water or other liquids.
Do not clean with direct water or steam jet or compressed air.
X Clean the Filling Chamber with a cotton swab soaked in isopropyl
alcohol.
6.3.2 Removing the Screen
X If a Screen is clogged, it must be replaced.
X Press in the Screen, using the Filling Chamber Tool along the groove of
the Filling Chamber.
Ö The Screen bulges and detaches from the groove.
X Clean the Screen with rinsing water or isopropyl alcohol.
6.3.3 Installing the Screen
X Arch the cleaned or new Screen.
X Insert the Screen into the Filling Chamber with the curved side facing up-
wards.
X Press the Screen at into the provided groove with the aid of the Filling Cham-
ber Tool.