A-4
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC
61000-4-11
Immunity
test
0% Uτ
(100% dip in
) for 0.5
cycle
0% Uτ
(100% dip in
) for 1 cycle
70% Uτ
(30% dip in Uτ)
for 25/30
cycles
0% Uτ
(100% dip in
) for 5 s
0% Uτ
(100% dip in
) for 0.5
cycle
0% Uτ
(100% dip in
) for 1 cycle
70% Uτ
(30% dip in Uτ)
for 25/30
cycles
0% Uτ
(100% dip in
) for 5 s
IEC 60601
test level
Compliance
level
Electromagnetic
environment –
guidance
Mains power quality
should be that of a
typical commercial
or hospital
environment. If the
user of the
MIGHTY
+
MEDIC
Vaporizer requires
continued operation
during power mains
interruptions, it is
recommended that
the MIGHTY
+
MEDIC Vaporizer be
powered from an
uninterruptible
power supply.
Power
Frequency
Magnetic Field
(50/60 Hz)
IEC 61000-4-8
30 A/m 30 A/m
Power frequency
magnetic elds
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Note: Uτ is the a.c. mains voltage prior to application of the test level.
INFORMATION ON ELECTROMAGNETIC COMPATIBILITY