ProStar 430 Autosampler Control Manual - 16 - 03-914947-33:2
Some of the following parameters depend on the injection
mode selected.
The Flush volume parameter is the amount of sample used to
fill the needle plus tubing connection to the injection valve (not
the loopfill volume). Only used for Full loop and Partial
loopfill injections.
NOTE: Flush volumes smaller than twice the volume of the needle plus tubing between
needle and injection valve may result in a decrease of performance.
NOTE: For more details about the injection modes, refer to the section Injection Mode at
the end of this manual.
Wash parameter group:
Select the Wash mode in the scrolling list. Three choices are
available: None, After injection (single run) or After sequence
(of injections).
If After injection is selected, enter the Wash volume in the
corresponding field.
If After sequence is selected, the wash procedure is fixed by
the instrument and the wash volume cannot be set.
Options parameter group:
If a plate cooling has been previously declared in the Galaxie
Configuration Manager, the Plate cooling fields can be
activated. Check Plate cooling to set the temperature setpoint
(4 to 40°C).
If Plate cooling automatic stop is checked, the temperature
control is switched off, then some variations of temperature will
be able to occur during a single run or a sequence.
If this option is not checked, the temperature between the
injections will be kept.