10. Troubleshooting
vatech A9 (Model name: PHT-30CSS) User Manual 121
English
Tail
code
Description
009
Appears when the inverter output current is higher than 1A during X-ray
irradiation (In EP, IP condition)
010
Appears when the inverter board falsely recognizes the exposure switch
signal as OFF after the irradiation On command
011
Appears when the X-ray OFF command is not sent to the inverter board
in 0.5 seconds after turning off the exposure switch
012
Appears when kV feedback is over -20kV compared to the setting value
during X-ray irradiation
013
Appears when kV feedback is over +20kV compared to the setting value
during X-ray irradiation.
014
Appears when the mA feedback value is less than 50% compared to
setting conditions during X-ray irradiation.
015
Appears when the mA feedback value is higher than 150% compared to
setting conditions during X-ray irradiation.
10.2.1.2 Subcode - Motor related error (002)
Tail
code
Description
021 Appears during rotator-axis motor origin movement
027 Appears during Cephalo sensor motor origin movement
037 Appears during generator tilting
10.2.1.3 Subcode – Exposure switch related error (003)
Tail
code
Description
060
It appears if the exposure switch is pressed when turning on the
equipment.
061
It appears if X-ray exposure is stopped by releasing the exposure
switch.
10.2.1.4 Subcode – Other error (004)
Tail
code
Description
102 Appears when there is no response during CAN communication.

Table of Contents

Other manuals for Vatech A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech A9 and is the answer not in the manual?

Vatech A9 Specifications

General IconGeneral
BrandVatech
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals