9
3 Product description
VEGAFLEX 82 • 4 … 20 mA/HART - two-wire
44222-EN-170220
•
Download the smartphone app "VEGATools" from the "Apple App
Store" or the "GooglePlayStore"
•
Scan the Data Matrix code on the type label of the instrument or
•
Enter the serial number manually in the app
This operating instructions manual applies to the following instrument
versions:
•
Hardware from 1.0.0
•
Software from 1.3.0
•
DTM from version 1.67.2
The instrument and the electronics version can be determined via the
product code on the type label as well as on the electronics.
•
Standard electronics: Type FX80H.-SIL
The scope of delivery encompasses:
•
Sensor
•
Optional accessory
•
Optionally integrated Bluetooth module
•
Documentation
– Quick setup guide VEGAFLEX 82
– Safety Manual (SIL)
– Instructions for optional instrument features
– Ex-specic"Safety instructions" (with Ex versions)
– Ifnecessary,furthercerticates
•
DVD "DTMCollection", included therein
– PACTware
– DTM Collection
– Instrumentmasterles(GSD)forProbusPA
– FDTcerticates
Information:
In this operating instructions manual, the optional instrument features
are described. The respective scope of delivery results from the order
specication.
3.2 Principle of operation
The VEGAFLEX 82 is a level sensor with cable or rod probe for con-
tinuous level measurement, suitable for applications in bulk solids.
DuetothequalicationuptoSIL2orhomogeneousredundantupto
SIL3 (IEC 61508) the VEGAFLEX 82 is suitable for the use in safety-
instrumented systems (SIS).
The safety function (SIF) can be a monitoring of the max. or min. level
or a combination of both.
High frequency microwave pulses are guided along a steel cable or
a rod. Upon reaching the product surface, the microwave pulses are
reected.Therunningtimeisevaluatedbytheinstrumentandoutput-
ted as level.
Scope of this operating
instructions manual
Versions
Scope of deliver
y
Application area
Functional principle -
level measurement