VEGAPULS 51K … 54K 5
21750-EN-031222
Reflected radar power dependent on the dielectric
constant of the measured product
2
0
0
10
25 %
40 %
5
10
5 %
20
30
40
50
%
4 6 8 12 14 16 18
20
ε
r
Product description
All products which are electrically conductive
reflect radar signals very well. Even slightly
conductive products provide a sufficiently
strong reflection for a reliable measurement.
All products with a dielectric constant ε
r
greater than 2.0 reflect radar pulses suffi-
ciently (note: air has a dielectric constant ε
r
of
1). Signal reflectivity grows stronger with
increasing conductivity or increasing dielec-
tric constant of the product. Hence, nearly all
substances can be measured.
Temperature influence: Temperature error absolutely
zero (e.g. at 500°C 0.018 %)
100 500 1000 1300 ˚C
0
0
0,01
0,02
0,03
%
0,018 %
0,023 %
Pressure influence: Error with pressure increase very
low (e.g. at 50 bar 1.44 %)
10
0
0
5
%
1,44 %
2,8 %
10
50
20 30 40 60
100
70 80 90 110 120 130 140
bar
0,29 %
3,89 %
With standard flanges of DN 50 to DN 250,
ANSI 2“ up to ANSI 10“ or G 1½ A and 1½“
NPT, the sensor antenna systems can be
adapted to various products and measuring
environments.
The high-quality materials can also withstand
extreme chemical and physical conditions.
The sensors deliver stable, reproducible
analogue or digital level signals with reliability
and precision and have a long useful life.
Continuous and accurate
Unaffected by temperature, pressure and
atmosphere content, VEGAPULS radar sen-
sors measure quickly, accurately and con-
tinuously the levels of widely varying
products.
VEGAPULS 50 sensors allow radar level
measurement in plants where it was hitherto
unthinkable because of high costs.
1.2 Application features
Applications
level measurement of any liquid, limited in
solids.
measurement also in vacuum
all slightly conductive materials and all
substances with a dielectric constant > 2.0
can be measured
measuring range 0 … 20 m.
Two-wire technology
power supply and output signal on one
two-wire cable (Loop powered)
4 20 mA output signal or HART
®
output
signal.
Rugged and abrasionproof
non-contact
high-resistance materials
Exact and reliable
accuracy 1 mm.
unaffected by noise, vapours, dusts, gas
compositions and inert gas stratification
unaffected by varying density and tem-
perature of the medium
measurement in pressures up to 40 bar
and product temperatures up to 200 °C.