EasyManua.ls Logo

Vega VEGAPULS 53K - 5.4 HART Handheld Adjustment

Default Icon
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48 VEGAPULS 51K … 54K
21750-EN-031222
Connection to a VEGA signal conditioning
instrument
If you operate a HART
®
compatible sensor on
a VEGA signal conditioning instrument, you
have to connect the sensor via a resistor
(see following table) during HART
®
adjust-
ment. This resistor, together with the internal
resistance of the instruments, provides the
value of 250 Ohm required for the HART
®
instrument. An inherent system load resist-
ance allows a corresponding reduction of Rx.
VEGA signal conditioning instr. Rx
VEGAMET 513, 514, 515, 602 50 … 100 Ohm
VEGAMET 614 no additional
VEGADIS 371 resistor
required
VEGAMET 601 200 … 250 Ohm
VEGASEL 643 150 … 200 Ohm
VEGAMET 513 S4, 514 S4
515 S4, VEGALOG EA card 100 … 150 Ohm
Set-up
Just connect the HART
®
handheld to the
signal cable, after having connected the
sensor to power supply.
Note:
If the resistance of the power supply is less
than 250 Ohm, a resistor must be connected
into the signal/connection loop during adjust-
ment.
The digital adjustment and communication
signals would otherwise be short-circuited
due to insufficient resistance of the supply
current source or the processing system,
and as a result, communication with the sen-
sor would not be ensured.
Ri > 250
+
-
250
Ri < 250
+
-
With any HART
®
handheld you can set up the
VEGAPULS series 50K radar sensors like all
other HART
®
compatible sensors. A special
DDD (Data Device Description) is not neces-
sary.
5.4 Adjustment with HART
®
handheld

Related product manuals