83
11 Supplement
VEGAPULS 62 • Probus PA
36505-EN-181127
Pressure Withoutreuxvalve Withreuxvalve
0.7 bar (10.15 psig) 3.4 m
3
/h 1.7 m
3
/h
0.8 bar (11.60 psig) 3.5 m
3
/h 1.9 m
3
/h
0.9 bar (13.05 psig) 3.6 m
3
/h 2.0 m
3
/h
1 bar (14.5 psig) 3.8 m
3
/h 2.2 m
3
/h
1.5 bar (21.76 psig) 4.3 m
3
/h 3.5 m
3
/h
2 bar (29.0 psig) 4.8 m
3
/h 4.0 m
3
/h
Thread G⅛
Closure with
Ʋ Non-Ex Dust protection cover of PE
Ʋ Ex Threaded plug of 316Ti
Reux valve - unmounted (as option with non-Ex version, included in the scope of delivery with Ex
version)
Ʋ Material 316Ti
Ʋ Seal FKM (SHS FPM 70C3 GLT), FFKM (Kalrez 6375)
Ʋ for tube diameter 6 mm
Ʋ Opening pressure 0.5 bar (7.25 psig)
Ʋ Nominal pressure stage PN 250
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable entry M20 x 1.5; ½ NPT
Ʋ Cable gland M20 x 1.5; ½ NPT (cable ø see below table)
Ʋ Blind plug M20 x 1.5; ½ NPT
Ʋ Closing cap ½ NPT
Material ca-
ble gland
Material seal
insert
Cable diameter
4.5 … 8.5 mm 5 … 9 mm 6 … 12 mm 7 … 12 mm 10 … 14 mm
PA NBR – ● ● – ●
Brass, nickel-
plated
NBR
● ● ● – –
Stainless
steel
NBR
– ● ● – ●
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, stranded wire 0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ Stranded wire with end sleeve 0.2 … 1.5 mm² (AWG 24 … 16)
Electromechanical data - version IP 66/IP 68 (1 bar)
Options of the cable entry
Ʋ Cable gland with integrated connec-
tion cable
M20 x 1.5 (cable: ø 5 … 9 mm)